# Before Your NHS Pilot: Who Owns the Clinical Safety Evidence?

> Who prepares, reviews and accepts clinical safety evidence before an NHS pilot? An owner matrix for manufacturer and deployer, plus open hazards.

Source: https://grandrounds.tech/insights/nhs-pilot-clinical-safety-evidence · Published 2026-11-23 · CTO Grand Rounds

A care organization agrees to a pilot, then asks for clinical safety evidence. The founder's team answers with a security pack, a feature list or a half-finished form. Each may help, but none says who is accountable for the safety case. Settle a narrower question first: who produces, who reviews and who accepts which evidence before the pilot starts?

**Short answer.** Name two parties first. The manufacturer builds and supplies the software; the deployer is the care organization that puts it into use. NHS standards pages describe DCB0129 as helping manufacturers of health IT software evidence clinical safety, and DCB0160 as addressed to health organizations responsible for clinical safety when they deploy and use IT systems (read on 18 November 2026). As illustrative planning guidance, not a rule from either standard, each party prepares and accepts its own evidence and a clinical-safety reviewer checks both. A CTO supplies technical evidence. A CTO cannot grant NHS approval or accept clinical safety alone, and an open mitigation stays open before pilot approval.

This page is for HealthTech founders and technical leads with a product and a planned NHS or care deployment. It does not classify a product as a medical device, interpret the standards or replace a clinical safety officer. Use it to draft an owner matrix, then take the matrix to the people who can accept each row. For the integration boundary, see [How Should a Clinical AI Documentation Product Connect to the EHR?](https://grandrounds.tech/insights/clinical-ai-ehr-integration)

## Which party owns the clinical safety evidence for this pilot?

Start from the real product and workflow, not from a checklist. Before you fill in the matrix, gather:

- a one-paragraph description of what the product does in the care workflow and what it does not do;
- the systems it reads from or writes to, and who operates each;
- any hazard log, safety case or test record you hold, with dates;
- the buyer's written requests: which documents, asked for by whom;
- the people who could accept each row. Where you have no name, write "unknown".

Then add one row per evidence item. This table is illustrative: an invented product and organization, no client result. The invented product shows a clinician a summary of a patient's GP record inside a care-coordination app.

| Evidence item | Prepared by | Reviewed by | Accepted by | Status |
| --- | --- | --- | --- | --- |
| 1. Intended use and product boundary | Supplier product owner, with CTO | Supplier clinical safety officer | Supplier clinical safety officer | Draft; wording not agreed with buyer |
| 2. Hazard log for the product | Supplier clinical safety officer, with engineering | Clinical-safety reviewer (not yet appointed) | Supplier clinical safety officer | In progress; hazards open |
| 3. Test evidence for each mitigation | Supplier engineering | Clinical-safety reviewer | Supplier clinical safety officer | One test not yet run |
| 4. Safety case report issued to the buyer | Supplier clinical safety officer | Care organization's clinical safety lead reads it | Supplier clinical safety officer issues it | Blocked by rows 2 and 3 |
| 5. Local deployment risk assessment | Care organization's clinical safety lead, with supplier input | Clinical-safety reviewer | Care organization's accountable owner (unknown) | Not started |
| 6. Local setup, training and go-live checks | Care organization's operations lead | Care organization's clinical safety lead | Care organization's accountable owner (unknown) | Not started |

Role titles are placeholders. The allocation and the "clinical-safety reviewer" lane are illustrative planning guidance for this example, not roles or duties taken from either standard. "Supplier clinical safety officer" means whoever the supplier has nominated; the buyer and the current standards decide what that role requires.

Read the table by its verbs. Preparing evidence is team work; accepting it is a decision for someone with the authority and clinical standing to make it. Here the CTO appears only in the prepare column: the CTO builds the tests and records, and others decide whether the risk is acceptable. Two cells say "unknown". Those are findings, and each needs a name before the pilot date means anything.

*Figure: Swimlane: manufacturer and deployer each prepare and accept their evidence; a reviewer checks both. — The same ownership as a swimlane. Source: original diagram; illustrative, not a compliance route.*

### Follow one hazard through the matrix

Take one invented hazard: a clinician switches patient in the GP system while the app still shows the previous patient's summary. Record it as one handoff.

| Field | Illustrative entry |
| --- | --- |
| Hazard | H-07: wrong patient context shown after a patient switch |
| Consequence and severity | Set by the clinical safety professionals; engineering does not rate it |
| Supplier mitigation | The app re-reads the patient identifier on each refresh and hides the summary if it no longer matches |
| Supporting test | T-14: switch patient mid-session in the integration test environment. Status: not run |
| Care organization's acceptance owner | Its clinical safety lead (name unknown) |
| Status | Open: mitigation designed, test not run, residual risk not accepted |

The CTO can close T-14 by running it and attaching the result. The CTO cannot close H-07. The hazard closes only when its owners accept it, and until then the pilot item stays open. This is an illustrative planning handoff, not a compliance verdict.

### What adjacent NHS integration experience shows

Unpublished profile text supplied to Grand Rounds lists Richard Sloggett as a panel CTO and says they were the sole technical contributor on a healthcare platform deployed in 200+ GP surgeries with NHS integrations (EMIS, SystmOne). That profile text is the only source here, and it describes technical work. It does not say who held clinical safety roles on that platform and it does not show safety evidence for your product. Use it to prompt a technical question: what boundary did the integration cross, and who accepted what? A separate profile item covers AI-assisted proposal and research workflows at a market access consultancy. That is not a clinical deployment and carries no weight here. Neither item establishes clinical safety clearance, and the profile does not state availability as a clinical safety officer.

## Separate DCB0129, DCB0160 and DTAC responsibilities

Keep three questions apart, because one document rarely answers all of them.

| Question | What this page can say, with source | Who to ask |
| --- | --- | --- |
| [DCB0129](https://standards.nhs.uk/published-standards/clinical-risk-management-its-application-in-the-manufacture-of-health-it-systems) | Listed as Active (reference DCB0129 Amd 24/2018); helps manufacturers of health IT software evidence clinical safety (read 18 November 2026) | The supplier's clinical safety officer and the current standard text |
| [DCB0160](https://standards.nhs.uk/published-standards/clinical-risk-management-its-application-in-the-deployment-and-use-of-health-it-systems) | Listed as Active (reference DCB0160 Amd 25/2018); addressed to health organizations responsible for clinical safety in deploying and using IT systems (read 18 November 2026) | The care organization's clinical safety lead |
| DTAC | No claim made on this page | The buyer's written request and the official criteria |
| Who qualifies for a safety officer role | No claim made on this page | The buyer and the current standard text |

Two cautions follow.

First, the standards are under review. [NHS England's consultation](https://www.engage.england.nhs.uk/patient-safety/national-review-of-clinical-risk-management-standa/) on DCB0129 and DCB0160 opened on 29 June 2026 and is listed as closed on 11 September 2026. The page says the review must reflect technologies such as AI supporting clinical decisions, interconnected systems sharing data across organizations, voice recognition and machine learning. The Safety-Critical Systems Club's [notice of the consultation](https://scsc.uk/club-news/public-consultation-opens-on-nhs-englands-clinical-risk-management-standards-dcb0129-0160) calls them "the mandatory standards that protect patients"; that is the Club's wording and the notice does not say who is bound. The standards pages above showed the 2018 references and no review outcome. Ask the buyer which versions and templates it applies today.

Second, do not treat DTAC as the DCB rows, or the reverse. This page makes no statement about DTAC criteria. Ask the buyer in writing which documents it expects, read the official criteria yourself, and add each requested item to the matrix as its own row with an owner.

## Map the product boundary, hazards and safety ownership

A hazard needs a place to attach. Four steps:

1. **Draw the boundary.** List every point where information enters or leaves the product: the record it reads, the screen where a clinician sees it, anything it writes back. Name who operates each point.
2. **List hazards at each crossing.** Your candidates are prompts for the clinical safety professionals, who decide what counts. Typical prompts: wrong patient context, stale or missing data shown as complete, and an unclear line between recorded information and generated content in an AI feature.
3. **Assign a control and an owner.** Software controls sit on the supplier side. A control that depends on local setup or staff behavior sits on the care organization's side, and the supplier says so in writing. A mitigation that exists only in a supplier's document may protect no one locally.
4. **Attach evidence.** For each control, name the test or record, its date, the build it covers and who reviewed it.

If the product's purpose has shifted from what the first hazard analysis assumed, for example by adding clinical advice, settle the intended-purpose wording first. This page does not decide medical-device classification.

### The handoff to the clinical-safety professional

A fixed shape helps the reviewer. Send:

- the product and intended use as currently worded, marked "unconfirmed" where it is;
- the boundary map and the operator of each crossing;
- the hazard list with status: open, mitigated but untested, or mitigated and tested;
- the evidence list with dates and build versions;
- rows with no owner, listed first;
- three questions: which standard versions the buyer applies, who accepts residual risk on each side, and what the buyer expects to see before the pilot starts.

## Check the safety evidence for the actual deployment

Evidence about a product is not evidence about a deployment. Test each row against the real pilot:

| Check | Question | If the answer is unknown |
| --- | --- | --- |
| Version | Does the evidence cover the build that will run in the pilot? | Re-date it, or mark the row stale |
| Route | Does it cover the actual integration route and environment? | Run the test in the pilot environment |
| Users | Does it cover the actual users and workflow? | Ask the care organization to describe the workflow in writing |
| Test | Did the supporting test run, and where? | Keep the mitigation open |
| Acceptance | Who accepted it, in what role, on what date? | Leave the row unaccepted |

Label each cell with the strongest status a document supports: documented, tested, accepted by a named owner, or unknown. "We believe so" is unknown, and a cell is not accepted because someone has seen it. Re-run the matrix when the intended use, integration route, user group or underlying model changes; in this example, any of those makes earlier evidence a question again.

Plan for one exception: the supplier has no available clinical safety officer. Row 2 then has no acceptor and row 4 cannot be issued. The honest status is "blocked", with the missing role named and the next step set: appoint or engage a suitable professional. A CTO who writes "complete" over that gap changes the record, not the risk.

## Work with the clinical and technical owners without claiming NHS approval

- **The CTO or technical lead** keeps the boundary map, versioned test records, release notes and an incident log, and answers "what does the product do, in which build, and how do we know?"
- **The supplier's clinical safety officer** (in this illustrative allocation) owns the manufacturer's hazard log and decides whether that evidence is accepted.
- **The care organization's clinical safety lead** (same caveat) owns the deployment assessment and its acceptance decision.
- **The sponsor or service owner** decides whether the pilot proceeds, using their advice.

A buyer reads a status line as a statement of fact, so write what the record shows:

| Avoid | Write instead |
| --- | --- |
| "NHS approved" | "Hazard log v0.3 prepared; review pending" |
| "DTAC passed" | "Buyer's DTAC request received; items 4 and 7 have no owner" (invented) |
| "Compliant" | "Safety case report drafted, not issued; blocked by T-14" |
| "Clinically safe" | "H-07 open; supplier mitigation designed; care organization acceptance pending" |

Grand Rounds does not stand in for those owners. Its [CTO for Healthtech & MedTech](https://grandrounds.tech/industries/healthtech) page, read on 18 November 2026, says it provides no clinical validation, medical review, regulatory or certification approval, or implementation or outcome guarantee, and that clinical and specialist acceptance stays with the responsible owners. If your gap is technical leadership, [Fractional CTO Services: What a Part-Time CTO Covers and When It Fits](https://grandrounds.tech/services/fractional-cto) covers the general role, and [Choosing a CTO for a Clinical AI Pilot Without Vague Compliance Claims](https://grandrounds.tech/insights/cto-for-clinical-ai-pilot) covers separating technical sign-off from clinical and quality or regulatory decisions. If the problem sits in the clinician's workflow, see [Is It the Model or the Handoff? Trace Where a Clinical AI Documentation Tool Stops Fitting](https://grandrounds.tech/insights/clinical-ai-workflow-fit).

## Scope the pilot and unresolved evidence before expansion

Let the matrix set the scope. Compare each row's status with the pilot you want to run:

| Row status | Pilot action in this example |
| --- | --- |
| Accepted by a named owner, covers the pilot build | The part of the pilot that relies on it can proceed, subject to the buyer's own decision |
| Open, outside the pilot scope | Record the exclusion in writing and keep the feature out |
| Open, inside the pilot scope | Hold that part of the pilot; name the owner and the next check |
| No owner | Name an owner before any date is set |

A smaller first pilot is often easier to evidence: fewer integration routes, one user group, one clinical setting. Treat a second site, another system or new functionality as a new run of the matrix, because the boundary and hazards may differ.

If the pilot is an AI workflow and the choice to proceed, repair, narrow or stop is unclear, [AI Readiness Assessment & Strategy](https://grandrounds.tech/services/ai-readiness) describes a separately agreed assessment of one workflow. That page, read on 18 November 2026, says the assessment is not an audit certification and that release needs its own evidence and approvals. It would not replace the clinical-safety acceptance in this matrix.

## Boundaries

All names, identifiers and hazards in the examples are invented to show the method. Statements about the standards reflect the pages linked above as read on 18 November 2026; the standards are under review and the full standard texts were not read for this page, so confirm current requirements and versions with the official standard and the buyer. This page is not clinical, legal or regulatory advice. It does not classify any product, and nothing in it states that a product is approved, cleared or compliant. Grand Rounds is an initiative of Dinarys, and a conversation may lead to cooperation with Dinarys.

## Next step

Copy the six-row matrix and fill it in for one pilot workflow. Mark every cell you cannot support with a document as "unknown" and put rows with no owner at the top. Send the matrix to the person at the buyer who owns clinical safety, and ask which rows they will review, which they will accept and which standard versions they apply. If the open rows are technical and you need leadership for them, start with [CTO for Healthtech & MedTech](https://grandrounds.tech/industries/healthtech). Any engagement is separately scoped, and none of it replaces the clinical-safety acceptance described above.